ВЛИЯНИЕ РЕЖИМОВ ЦИКЛИЧЕСКИХ НАГРУЗОК НА РАЗВИТИЕ ДЕФОРМАЦИЙ В ПОВРЕЖДЕННЫХ БЕДРЕННЫХ КОСТЯХ С СИСТЕМАМИ ФИКСАЦИИ ПЕРЕЛОМОВ

Authors

  • Н. С. Шидловский National Technical University of Ukraine «Kyiv Polytechnic Institute», Kyiv, Ukraine
  • Д. Е. Шпак National Technical University of Ukraine «Kyiv Polytechnic Institute», Kyiv, Ukraine
  • Д. В. Ковалёв National Technical University of Ukraine «Kyiv Polytechnic Institute», Kyiv, Ukraine

DOI:

https://doi.org/10.20535/2305-9001.2011.63.56036

Keywords:

The fixation stability of bone fragments to static and cyclic axial loadings were investigated

Abstract

This paper presents the biomechanical background of different fixation methods performed on simulated, unstable intertrochanteric hip fractures. There were used two types of fixation: by endoprosthesis pedicle and DHS plate. The fixation stability of bone fragments to static and cyclic axial loadings were investigated. The results demonstrated more high stiffness of fixation with the use of total hip cement arthroplasty combined with additional wire mesh, that is biomechanically superior to DHS plates’ fixation of unstable intertrochanteric hip fractures. The stiffness of fixation with the use of total hip cement arthroplasty is closed to stiffness of an uninjured bone under 400N loading.

Author Biographies

Н. С. Шидловский, National Technical University of Ukraine «Kyiv Polytechnic Institute», Kyiv

канд.техн.наук, доц.

Д. Е. Шпак, National Technical University of Ukraine «Kyiv Polytechnic Institute», Kyiv

доц., канд.техн.наук

Д. В. Ковалёв, National Technical University of Ukraine «Kyiv Polytechnic Institute», Kyiv

маг.

Published

2015-12-15

Issue

Section

Статті